Roche Accu-Chel Manuel de service Page 107

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 135
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 106
107
Symbols, problems and error messages
12.2 Problems
The meter will not turn on.
You have not removed the protective lm from the
battery.
Remove the lm from the battery (see Chapter
2.3).
The battery is empty or there is no battery in the
meter.
Insert a new battery.
You have inserted the battery the wrong way
round.
Remove the battery and place it in the battery
compartment as specied.
The ambient temperature is low and the battery is
almost empty.
Ensure that the ambient temperature is between
+10 and +40 °C and wait for the meter to warm
up. Insert a new battery.
The electronics have become damp through
condensation.
Allow the meter to dry slowly.
The meter is defective.
Please contact your customer support and service
centre.