Roche Accu-Chek Aviva Informations techniques Page 237

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 282
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 236
235
8
Display Shows Possible Cause(s) Possible Solution(s)
E-51 Bad Strip
Error
j Your blood glucose may be
extremely low.
If you see this error message after you applied blood
to the test strip:
j If you are experiencing any of the common
symptoms of low blood sugar, contact your
healthcare professional immediately.
j Treat your low blood sugar as recommended by
your healthcare professional.
j If this does not match how you feel, repeat the
blood glucose test and see Chapter 2, Testing Your
Blood Glucose.”
j The test strip is damaged.
j The test strip is not properly
inserted into the meter.
If you see this error message before you applied
blood to the test strip:
j Remove the test strip and reinsert it, or replace it if
damaged. If the message reappears, contact
Roche.
E-52 Code Key
Error
j The activation chip is
incorrect.
j Please contact Roche.