Display Shows Possible Cause(s) Possible Solution(s)
Pump Not Available
j
The pump is out of the
communication range of the meter.
j
Ensure the meter and pump have
Bluetooth wireless technology turned on.
j
Ensure the meter and pump are within
communication range.
j I
f the low battery icon is displayed on the
meter, the meter cannot communicate
with the pump. Replace the batteries.
Time Mismatch
j
The time or date on the meter is
more than ve minutes different
than the time or date on the pump.
j
The meter time and date were changed to
match the pump time. Ensure the time
and date are correct. If the time or date
are not correct, change the time and date
on the pump.