Roche Accu-chek Performa Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Non Roche Accu-chek Performa. Performa ® Meter Information - Vancouver Island Health Authority Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer

Résumé du contenu

Page 1 - Blood Glucose Meter Training

Blood Glucose Meter Training1

Page 2 - Disclaimer

Accu-Chek Performa Operator Training © VIHA 2013 Coding a meter1. Make sure the meter is OFF2. Turn the meter over3. Remove the old code chi

Page 3 - What is changing?

Accu-Chek Performa Operator Training © VIHA 2013 Coding a meter cont’d4. Turn the code chip over so the number faces down5. Push chip into t

Page 4 - Learning Outcomes

Accu-Chek Performa Operator Training © VIHA 2013 Code Chip• Each box of test strips contains a code chip• Ensure this new chip is inserted i

Page 5

Accu-Chek Performa Operator Training © VIHA 2013 Cleaning & Disinfecting the MeterClean the meter after each client test with a Percept

Page 6 - • 5-second test time

Accu-Chek Performa Operator Training © VIHA 2013 Meter Cleaning Steps1. Put on clean gloves (and gown if required). 2. Make sure meter is of

Page 7 - Arrow Buttons

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Test Strips15

Page 8 - Changing the Battery

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Test Strip - Features• 0.6 µL sample size• Wipe 1stdrop and use 2nddrop•

Page 9 - Changing the Battery cont’d

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Control Solution• Only use Accu-Chek Performa Control Solutions• Write t

Page 10 - Coding a meter

Accu-Chek Performa Operator Training © VIHA 2013 Why do a Quality Control ?• Performing a Quality Control verifies the reliability of the te

Page 11 - Coding a meter cont’d

Accu-Chek Performa Operator Training © VIHA 2013 When to do Quality ControlA Quality Control should be done:• Immediately prior to client te

Page 12 - Code Chip

Accu-Chek Performa Operator Training © VIHA 2013 Disclaimer• This information is intended for use by Health Care Professionals • Patients wh

Page 13 - Remember:

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – for Healthcare Professionals 1. Insert the test strip into the

Page 14 - Meter Cleaning Steps

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d3. Select the control solution to be used4. Place the met

Page 15

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d6. Touch the second drop to the front yellow edge of the

Page 16 - • Wipe 1

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d– The QC result appears on the display, along with a cont

Page 17

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Control Test – cont’d10. Result is in range if “OK” & the result alternate

Page 18 - Why do a Quality Control ?

Accu-Chek Performa Operator Training © VIHA 2013 Troubleshooting - Quality Control is Out of RangeReview this list for possible causes:– Wer

Page 19 - When to do Quality Control

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test 1. Insert a test strip, the meter turns on 2. Make sure the

Page 20 - Professionals

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test – cont’d3. A test strip and flashing blood drop appear on t

Page 21 - 4. Place the meter on

Accu-Chek Performa Operator Training © VIHA 2013 Performing a Blood Glucose Test – cont’d5. Wipe the first drop away and touch the second dr

Page 22 - (means that enough sample

Accu-Chek Performa Operator Training © VIHA 2013 Hi and Low Results Display• Hi Level - greater than 33.3 mmol/L• Low Level – less than 0.6

Page 23

Accu-Chek Performa Operator Training © VIHA 2013 What is changing?Roche ACCU-CHEK® meters will be the meters used by all Health Authorities

Page 24

Accu-Chek Performa Operator Training © VIHA 2013 Reference RangesVIHA Glucose Meter Normal Reference Interval 4.0 - 11.0 mmol/L(The meter wi

Page 25

Accu-Chek Performa Operator Training © VIHA 2013 VIHA Critical Values(follow your facilities procedures for appropriate actions)Critical Hig

Page 26 - 2. Make sure the code number

Accu-Chek Performa Operator Training © VIHA 2013 Documenting ResultsRemember - Always document the result as soon as possible to avoid error

Page 27

Accu-Chek Performa Operator Training © VIHA 2013 Viewing Results - AveragesThe Perform meter has the ability to show 7, 14 or 30 day averag

Page 28

Accu-Chek Performa Operator Training © VIHA 2013 Limitations of ACCU-CHEK®Test Strips (Refer to test strip package insert for further inform

Page 29 - Hi and Low Results Display

Accu-Chek Performa Operator Training © VIHA 2013 Maintaining Competency Competency on the blood glucose meter is your professional responsi

Page 30 - Reference Ranges

Accu-Chek Performa Operator Training © VIHA 2013 Learning OutcomesThese slides should help you understand: • Basic meter features • How to p

Page 31 - Critical High Critical Low

Accu-Chek Performa Operator Training © VIHA 2013 Review of existing BGM practice/policy Client identification must be confirmed using a min

Page 32 - Documenting Results

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Meter Features• 5-second test time• 0.6 – 33.3 mmol result range• Large

Page 33 - Viewing Results - Averages

Accu-Chek Performa Operator Training © VIHA 2013 Accu-Chek Performa Meter FeaturesDisplayRight and Left Arrow ButtonsOn/Off/Set ButtonBatter

Page 34 - Key Points:

Accu-Chek Performa Operator Training © VIHA 2013 Changing the BatterySteps:1. Open the battery door on the back of the meter by pushing the

Page 35 - Maintaining Competency

Accu-Chek Performa Operator Training © VIHA 2013 Changing the Battery cont’d93. Insert the new battery with the + side up.4. Put the battery