Inform IIQuick Reference Guidefor Healthcare ProfessionalsBLOOD GLUCOSE MONITORING SYSTEM
10Meter, Front ViewACCU-CHEK Inform II MeterTest Strip PortTouchscreenOn/Off Button
11Meter, Back ViewACCU-CHEK Inform II MeterACCU-CHEK Inform II MeterBarcode Scanner WindowBattery PackReset ButtonInfrared Window
12ACCU-CHEK Inform II MeterThe base unit:•Chargesthemeterbatterypack.•Communicateswithadatamanagement system, if available.Base Unit
13Patient TestingNote: Prior to testing, ask patient to wash hands with soap and water, then let dry. If patient is unable to wash hands, wipe with al
144. Touch Patient Test to display the Patient ID screen.How to Perform Patient Testing
155. Enter or barcode scan patient ID, then touch . You are prompted to confirm the test strip lot.Patient TestingHow to Perform Patient Testing
166. Verify the test strip lot.•If correct, touch . You are prompted to insert the test strip.•If incorrect, touch to select a different test str
177. Slide the test strip into the test strip port as far as it will go in the direction indicated by the arrows on the test strip.8. Wait until the
189. Apply the drop of blood to the front edge (yellow dosing area) of the test strip. Do not apply the blood to the top of the strip.Blood is pulled
1910. When the test is completed, the result is displayed. Touch to enter the desired comments. Touch to return to the Main Menu.How to Perform Pat
2
2011. Remove the test strip and dispose of it in accordance with applicable regulations and directives for disposal of potentially infectious samples
21Control TestingHow to Perform Control TestingGlucose control tests should be run in the following circumstances:•Before using the meter for the pat
22How to Perform Control Testing•If questionable test results are displayed repeatedly•If you wish to test the performance of the systemIn addition,
231. Press the On/Off button to power on the meter.2. Once the Power Up screen appears, touch to proceed to the Operator ID screen, or wait 5 sec
243. Enter or barcode scan operator ID, then touch to display the MainMenu screen.4. Touch Control Test to display the Control Test screen.How to
255. Touch the control level to be run.How to Perform Control TestingControl Testing
266. Verify the control lot number.•If correct, touch . You are prompted to confirm the test strip lot.•If incorrect, touch to enter or barcode
277. Verify the test strip lot.•If correct, touch . You are prompted to insert the test strip.•If incorrect, touch to select a different test st
288. Slide the test strip into the test strip port as far as it will go in the direction indicated by the arrows on the test strip.9. Wait until the
2910. Apply a drop of glucose control solution to the front edge (yellow dosing area) of the test strip. Do not apply the control solution to the top
3Table of ContentsImportant ... 4General Information ... 6ACCU-CHEK Inform II Met
3011. When the test is completed, the result is displayed. Touch to enter the desired comments. Touch to continue with the next level, if necessary
3112. Remove the test strip and dispose of it in accordance with applicable regulations and directives.How to Perform Control TestingControl Testing
321. Press the On/Off button to power on the meter.2. Once the Power Up screen appears, touch to proceed to the Operator ID screen, or wait 5 sec
33Storing Test Strip Lot Information4. Touch Strip Lots to open the related menu.5. Touch Add if you want to add the information for a new test strip
346. Insert the new code key in the opening of the code key reader. An LED starts flashing green and signals that the reader is ready to transfer dat
357. Hold the meter 4-6 in above the code key reader so that a connection can be made between the two infrared windows.8. Touch to begin downloadin
369. Confirm the test strip lot information.•Touch to store the data for this lot number without changes, or•Touch to modify the data for this te
37The meter does not power on.Solution: Confirm that the meter battery pack is charged correctly.The “QC Due: Immediately” message appears on the scre
05234654001 2012-10 USAACCU-CHEK, ACCU-CHEK INFORM, and COBAS are trademarks of Roche.©2012 Roche Diagnostics. All rights reserved. Roche Diagnostic
4WARNING: Protection against infection• Refer to the operator's manual for the proper cleaning and disinfection instructions for this system.•
5WARNING: Protection against infection (continued)• Do not dispose of the battery pack with normal domestic waste. As a component of the ACCU-CHEK In
6General Information• Refer to the ACCU-CHEK Inform II System Operator’s Manual and package inserts for detailed information. Be sure to read the “I
7General InformationYour system is ready to use when it has been charged in the base unit.Meters can be configured according to your institution’s nee
8QC Lockout occurs when •patient testing is attempted and controls have not been run in the time interval or frequency designated by your institutio
9ReagentsYou need the following reagents to perform patient tests and glucose control tests:•ACCU-CHEK Inform II Test Strips•ACCU-CHEK Inform II Con
Commentaires sur ces manuels