92
Now check that the displayed value is within the permitted range.
Examine the test strip container and locate the “Accu-Chek Active
Control” table.
The table has two rows listed as “1” and “2”, as well as two
columns giving ranges in mg/dL and mmol/L.
A
CCU
-C
HEK
®
Active Control
mg/dL mmol/L
1 36 – 66 2,0 – 3,7
2 131 – 177 7 ,3 – 9,8
씰 If you performed the test with Accu-Chek Active Control 1,
refer to row 1 for the permitted range.
씰 If you performed the test with Accu-Chek Active Control 2,
refer to row 2 for the permitted range.
If the result is within the stated range, you need to still carry out a
visual plausibility test. It is important that this check be performed
within 30–60 seconds after control solution was applied. Any later
than this and comparison is no longer possible owing to excessive
discolouration of the test strip.
If the result is outside the stated range, perform a second quality
control test. If the second result is still outside the range, please call
your customer support and service centre.
씰 Pull the strip out of the monitor. The result is saved as a control
reading (which is ignored when the averages are calculated),
and the monitor switches off.
93
6.3 Quality control test procedure
Values are for illustration only.
015134 AccuChek Active_Eng.indd 92-93 12.08.2004, 14:54:38