118
Measurement optics error. Clean the measurement
optics (Chapter 7.1). Check that the test strip is
lying flat and straight in the test strip guide. If the
message persists, the monitor has a fault. Please
call your customer support and service centre.
Although rarely occuring - you applied too little
blood to the strip. Repeat the test with a new
Accu-Chek Active test strip.
The measurement optics are dirty or you have
inserted a test strip that has already been used
or is expired. Clean the measurement optics
(Chapter 7.1) and repeat the test with a new
Accu-Chek Active test strip.
The monitor was exposed to a strong electro-
magnetic field. Move somewhere else.
The coding chip was removed during testing, has
a fault, or is not properly inserted in the monitor.
Remove the coding chip and re-insert it.
The test strip was not fully inserted, was moved
during testing, or was dosed with blood too soon.
Repeat the test with a new test strip and ensure
that it snaps into place.
The coding chip is missing or has a fault.
The Accu-Chek Active monitor has a fault and
cannot be used. Please call your customer support
and service centre.
You applied too little blood to the test strip, or the
test strip was not lying flat and straight in the test
strip guide. Repeat the test with a new test strip.
The ambient temperature or the temperature of
the monitor is outside the permitted range. Move
the monitor to a room-temperature environment
and wait until the monitor has acclimatised and the
error message has disappeared.
119
8.3 Troubleshooting
015134 AccuChek Active_Eng.indd 118-119 12.08.2004, 14:54:52