Testing a capillary blood sample
64
Test
09:15 am
09/16/2009
P-ID: PID01
Code: 184
Smith, Henry
Op: Operator 1
Par: PT
The hourglass symbol shows that the test
strip is warming up. When the warming-
up process is complete, a further beep
tone (provided the beeper is turned on)
indicates that you can now apply blood.
Test
09:15 am
09/16/2009
P-ID: PID01
Code: 184
180 SEC
Smith, Henry
Op: Operator 1
Par: PT
The blood drop symbol flashes to indicate
that the meter is ready to perform the test
and is waiting for blood to be applied.
At the same time, a 180-second
countdown begins. You must apply the
drop of blood to the test strip within this
time; otherwise, you will receive an error
message.