Quality control
74
QC Test
09:15 am
09/16/2009
184
Op: Operator 1
QC Test
09:15 am
09/16/2009
QC 248
QC 379
QC 462
New
QC 332
*
QC Test
09:15 am
09/16/2009
Code: 184
Op: Operator 1
Par: PT
QC
19 Select the code already stored for
your current control solution, or touch
New to use a new control solution.
Please note: When you first run your
control, the QC TEST screen will not
display. This screen will display the
next time you use the control.
If you are using a new test strip lot and
have not inserted the test strip code chip
yet, you must do so now. As with the test
strips, a quality control code chip is also
provided with the liquid quality controls.
This chip tells the meter the acceptable
ranges of results for that lot of controls.
The information on the code chip is
retained in the memory so you can use
the same control solutions at any time.
If you are using a new control solution,
remove the code chip from the meter and
insert the code chip that came with the
control solution instead.
If the code chips get mixed up, here’s
how to tell them apart: The name on the
code chip that came with the test strips
starts with the letter S, and the name of
the code chip that came with the control
solution starts with the letter C.