Roche CoaguChek XS System Manuel d'utilisateur Page 91

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 108
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 90
Error messages
91
Error Description/Solution
Error: Cover
The test strip guide cover is not properly closed.
Error
Solution
Close the test strip guide cover.
The error message disappears. The meter is ready.
Error: Test strip
The test strip is unusable or the test strip is not a CoaguChek
XS PT test strip.
Error
Solution
Remove the test strip and repeat the test with a new or
appropriate test strip.
Error: Code chip
The code chip is missing, not properly inserted in the meter,
or damaged.
Error
JVKL
Solution
Insert the code chip, or remove it and re-insert it.
Error: Test strip
The test strip is beyond its expiration date.
Error
EXP
Solution
First check whether the date setting is correct in the meter.
If it is not, set the correct date.
If the date is correct: Turn the meter off, remove the code
chip and test strip. Use a test strip from a new lot of test
strips and insert the code chip that came with the new lot.
Error: Test strip
The test strip is unusable.
Error
4
Solution
Turn the meter off, remove the test strip and re-insert it. If the
error message appears again, discard the unusable test strip
and use a new one.
Troubleshooting error messages when preparing to test
Vue de la page 90
1 2 ... 86 87 88 89 90 91 92 93 94 95 96 ... 107 108

Commentaires sur ces manuels

Pas de commentaire