Initial Startup
109
On the second settings screen, you can select options for
handling lot numbers with glucose control and linearity
tests.
1 Touch the desired option to enable it:
Control Lot Verification
Display only: The lot number is shown on-screen
but the operator cannot confirm it or select a
different lot number.
Confirmation: The operator must confirm the lot
number displayed and can enter alternative lot
numbers manually or via barcode scanner.
Lot entry: The operator must enter the lot number.
Scan only: Lot numbers can be verified via barcode
scanner only.
Linearity Lot Verification
Display only: The lot number is shown on-screen
but the operator cannot confirm it or select a
different lot number.
Confirmation: The operator must confirm the lot
number displayed and can enter alternative lot
numbers manually or via barcode scanner.
Lot entry: The operator must enter the lot number.
2 Touch to save the settings (for both settings
screens), or
3 Touch to return to the first settings screen.
Reagent Options
03/16/10
12:48pm
Scan only
Display only
Conrmation
Display only
Conrmation
Lot entry
Lot entry
Control Lot Verication:
Linearity Lot Verication: