Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
64
Information about the expiration date and parameters for
control solutions is subsequently displayed.
1 Touch to store the data for this lot number in
the meter without changes, or
2 Touch to modify the data for this lot number
before storing it in the meter.
Editing test strip data As mentioned at the beginning of this chapter, you can
edit several parameters for test strips, the expiration date
and the value ranges for control solutions.
3 Use the keypad to enter the desired expiration date
(use two digits and leading zero, if necessary). It is
not possible to enter a date beyond the expiration
date stored in the key.
4 Touch to accept the modified date and
continue to the value ranges.
Add Strip Lot
03/16/10
12:48pm
Strip Lot Conrmation
D-530: Do you want to
use the suggested
values for strip 545603?
Use by: 02/10/11
L1(Lo): 2.4-4.1 mmol/L
L2(Hi): 16-21.6 mmol/L
Add Strip Lot
03/16/10
12:48pm
Strip Lot Conrmation
D-530: Do you want to
use the suggested
values for strip 545603?
Use by: 02/10/11
L1(Lo): 2.4-4.1 mmol/L
L2(Hi): 16-21.6 mmol/L
123
456
789
0
Strip Expiration 12:48
pm
03/16/10
07/31/10