Linearity Testing
81
In the Linearity Test menu, the levels available for the lin-
earity test are displayed.
5 Touch L1 to start the subsequent test with this
(first) level.
Inserting test strips After selecting the level, you are prompted to insert the
test strip.
1 Remove the test strip from the test strip vial and
close the vial again with the cap.
2 Hold the test strip so the lettering “ACCU-CHEK” is
facing upward.
3 Slide the test strip into the test strip port as far as it
will go in the direction indicated by the arrows on
the test strip. The meter beeps.
Linearity Test 12:48pm
03/16/10
L1
L5
L3
L6
L4
L2
Linearity 12345678
Strip Lot 545794
Linearity Test 12:48pm
03/16/10
Linearity 12345678 L1
Strip Lot 545794