Troubleshooting
133
Base unit
LED is not illuminated Not connected to power supply unit or power supply unit is
defective, base unit is defective, or power outlet not active.
Disconnect and connect the power supply again.
If the error persists, contact your Roche representative.
LED flashes red Communication or configuration error.
Check the configuration and/or the connection to the Data
Management System (DMS). Contact the DMS or IT
administrator of your institution.
If the error persists, contact your Roche representative.
Automatic shutdown
The meter powers off after a configurable time without
activity (e.g., pressing a key, touching the screen) to
conserve energy. In addition, the meter may power off
automatically for the reasons listed below. Reactivate the
meter/screen as described in the following:
Shutdown after time specified by
system administrator (default is
10 minutes, configurable by system
administrator)
Press the On/Off button on the top of the meter.
Battery Low Charge the battery by placing the meter in the base unit.
Battery Critically Low Charge the battery by placing the meter in the base unit.
Check that the battery pack is correctly installed and
connected.
Replace the defective battery pack.
Follow the instructions in Chapter 9 to avoid data loss when
replacing the battery pack.