Roche ACCU-CHEK Inform II Manuel de l'opérateur Page 127

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 160
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 126
Maintenance and Care
127
Disinfecting the base unit
Remove the meter from the base unit prior to
disinfecting.
Unplug the base unit before disinfecting.
Be certain that solutions contain acceptable active
ingredients for disinfecting the system. If using
commercially available pre-moistened disinfecting
cloths, squeeze off excess disinfecting solution
or blot on a dry paper towel to remove any excess
disinfecting solution.
Wipe the base unit thoroughly. Do not wipe over the
cable connector areas in the rear of the base unit.
If solution enters any of these areas, dry them off
immediately using a soft, dry cloth.
Allow the base unit to air dry for the recommended
contact time according to the disinfecting solution
product labeling. Do not place the meter in the base
unit until all disinfecting steps are completed.
Dry the base unit thoroughly with a dry cloth or
gauze and visually verify that no solution is seen in
the connector and cable areas at the completion of
disinfecting. Ensure that the base unit
(including connectors) is thoroughly dried
after disinfecting.
Plug in the base unit.
NOTICE
Do not allow liquid to enter the connectors. If liquid does
get into the connectors, immediately dry the components
with a dry cloth or gauze. If solution is allowed to col-
lect in any base unit connector, severe damage can
occur.
NOTICE
Do not spray the base unit directly with solutions as this
could cause the solution to enter the case and damage
the electronic components.
Vue de la page 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 159 160

Commentaires sur ces manuels

Pas de commentaire