Glucose Control Testing
49
Information stored during
glucose control testing
The following information is stored for every glucose con-
trol test using control solution:
■ Glucose control test result
■ Lot number of the control solution
■ Operator ID (if configured)
■ Level of control solution (L1 or L2)
■ Lot number of the test strips
■ Time and date of test
■ Comments (if present)
■ Out of range measurements
Control solutions For blood glucose test strips, control solutions have two
levels:
■ Level 1 (L1): Lo (low values in test results)
■ Level 2 (L2): Hi (high values in test results)
Preparing to run a glucose
control test
Aside from special preparations (see the following sec-
tion), a glucose control test is run in the same manner as
a patient test:
■ At least one code file for test strips must be stored
in the meter and match the lot number of the test
strips used (see Chapter 6).
■ The proper test strips must be available.
■ An operator ID must be entered (with password, if
applicable), if the meter is configured for login.