Storing Test Strip, Control Solution, and Linearity Solution Information in the Meter
65
The parameters for control solutions consist of four sepa-
rate values.
5 Use the keypad to enter the desired values one
after another:
Minimum limit value for Level 1
Maximum limit value for Level 1
Minimum limit value for Level 2
Maximum limit value for Level 2
6 Touch to confirm each separate entry and
continue to the next entry.
123
456
789
0
Control L2 Max 12:48
pm
03/16/10
21.6 mmol/L
123
456
789
0
Control L2 Min 12:48
pm
03/16/10
16 mmol/L
123
456
789
0
Strip Expiration 12:48
pm
03/16/10
07/31/10