Maintenance and Care
126
WARNING
Do not spray directly into the strip port!
Moisture in the strip port may lead to incorrect blood
glucose results. If you suspect that moisture may have
entered the strip port, perform a glucose control test.
NOTICE
Do not allow liquid to enter the test strip port or allow
pooling of liquid on the touchscreen. If liquid does get into
the test strip port, immediately dry the components with a
dry cloth or gauze. If solution is allowed to collect in
any meter opening, severe damage can occur.