Initial Startup
112
Options for test strips
This menu allows you to select options for handling lot
numbers and specify whether the operator is allowed to
edit the expiration date and limit values on the meter
(optionally with password required).
1 Touch the desired option to enable it:
n Reagent Editing
– Allowed: Enables the operator to edit the expiration
date and limit values
– Password Needed: Enables the operator to edit the
expiration date and limit values only after entering
a password
n Strip Lot Verication
– Display only: The lot number is shown on-screen
but the operator cannot conrm it or select a
different lot number.
– Conrmation: The operator must conrm the lot
number displayed and can enter alternative lot
numbers manually or via barcode scanner.
– List selection: The operator can also select from a
list of stored lot numbers.
– Scan only: Lot numbers can be veried via barcode
scanner only.
2 Touch
to change to the second screen of
options, or
3 Touch
to exit this menu without saving any
changes.
When using a data management system for
conguration, it is possible to fully disable the
ReagentEditing option.