Glucose Control Testing
53
Information stored during
glucose control testing
The following information is stored for every glucose
con trol test using control solution:
n Glucose control test result
n Lot number of the control solution
n Operator ID (if congured)
n Level of control solution (L1 or L2)
n Lot number of the test strips
n Time and date of test
n Comments (if applicable)
n Out of range measurements
Control solutions
For blood glucose test strips, control solutions have two
levels:
n Level 1 (L1): Lo (low values in test results)
n Level 2 (L2): Hi (high values in test results)
Preparing to run a glucose
control test
Aside from special preparations (see the following
sec tion), a glucose control test is run in the same manner
as a patient test:
n At least one code le for test strips must be stored
in the meter and match the lot number of the test
strips used (see Chapter 6).
n The proper test strips must be available.
n An operator ID must be entered (with password, if
applicable), if the meter is congured for login.