Glucose Control Testing
61
Performing a STAT test
The meter can be congured to allow a STAT patient
glu cose test to be run even if the meter is in QC Lockout
or Download Lockout. This option is to be used in
situations with critical patients. The system administrator
can allow control tests to be delayed one to nine times.
Run STAT appears in the QC Lockout warning box under
the following conditions:
n You are successfully logged in and have selected
Patient Test from the main menu.
n A glucose control test is required (due to specied
control intervals or other conditions).
n The administrator has enabled the STAT test option
during setup.
n The number of STAT tests available has not been
exceeded.
If these conditions are fullled, two buttons appear in the
warning message that allow you to choose the next step:
n Touch Run QC to run the required control test
instead of a patient test.
n Touch Run STAT to perform a patient test even if a
glucose control test is required. The status as STAT
test is stored with the data record for the test.
If the meter is in a Download Lockout status, and no
STAT test is available, the meter can be unlocked by a
system administrator (see page 122).