General Product Information
145
A Appendix
Table of conguration options
This section provides an overview of all the settings
avail able.
The two rightmost columns describe the accessibility of a
conguration parameter on the device (Setup Menu) and
via the data management system (DMS).
“Y” (Yes) means, that this parameter is available, “N” (No)
means, that this parameter is not available using the
respective conguration method.
DEVICE: If the meter is congured by a system
administrator using a data management system, the
conguration options on the meter may be disabled to
avoid conicts in the settings. This option is indicated
by the use of parentheses e.g., (Y).
DMS: Conguration options may vary depending on
available DMS settings.
Conguration options
Subject/Attribute Range Default Device DMS
Timer
Shutdown timeout (meter powers off if no
activity)
30 - 3,600 s 300 N Y
Display: Low Power Mode after
(0 = feature disabled)
0 - 120 s 30 Y Y
Beeper
Beeper volume 0: low
1: medium
2: high
2 Y Y
Measurement Flow
Comments required 0: optional
1: out of range
2: required
3: disabled
0 N Y